ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 61

R Series Service Manual
9650-0903-01 Rev. J
2-9
NO QRS DETECT
Unit is in sync mode and heart rate is < 20 BPM or QRS
amplitude is too low for proper synchronization.
Increase ECG size and/or change lead.
NO SHOCK ADV.
No shock advised. Advisory message when analysis finds
non-shockable rhythm.
NOISY ECG
Number of noisy analysis intervals exceeds threshold.
Stop all patient movement. Check connections.
Press Analyze button again.
OPEN AIR DISCHARGE
Measured Defib Impedance was greater than 1000 ohms
and no energy was delivered.
PACER DISABLED
User prompt issued simultaneously with other pace faults if
pacing is disabled.
Replace high voltage module or system board.
PACER FAULT 115
PD Pace failure during POST.
Replace PD engine.
PACER FAULT 117
PD Pace functional safety failure while running.
Replace PD engine.
PACER FAULT 121
PD Pace failure while running.
Replace PD engine.
PACER FAULT 122
PD module Defib/Pace failure during POST.
Replace PD engine.
PACER WARNING 124
PD Pace warning while running.
Replace PD engine.
PADDLE FAULT
Cannot detect type of accessory attached to the universal
cable.
Replace paddles, internal paddles, system board,
high voltage module and/or universal cable.
PEDIATRIC PADS IN USE
The one of the Pediatric ID is missing.
PERFORM CPR
Message displayed upon entered CPR period specified by
AHA protocol.
POOR LEAD CONTACT
One or more ECG leads are poorly connected or not
connected to patient. (User configurable.)
Check electrode attachment to patient, cable
connector to electrode, and cable to unit
connector.
Error Message
Explanation
User
Advisory
Technical Action