Advisory Defibrillation Procedure
9650-0904-01 Rev. K
ZOLL R Series Operator’s Guide
5–3
Note:
If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.
The control buttons for Manual mode light up on the front panel and the unit displays DEFIB
120J SEL on the monitor.
Energy Select
The default energy selections for adult patients are:
•
Shock 1 - 120 joules
•
Shock 2 - 150 joules
•
Shock 3 - 200 joules
When used with OneStep Pediatric electrodes, the default energy selections for pediatric
patients are:
•
Shock 1 - 50 joules
•
Shock 2 - 70 joules
•
Shock 3 - 85 joules
WARNING!
Use only OneStep Pediatric electrodes to defibrillate patients under 8 years of age in
Advisory mode. Use of adult electrodes, or pediatric electrodes other than OneStep
Pediatric electrodes, can result in the delivery of excessive energy doses.
After the third shock, all subsequent shocks are delivered at the same energy as the third shock
in both Adult and Pediatric modes.
If medical protocols allow, you may select a different energy level using the energy select up
and down arrow buttons on the front panel. The new energy setting displays on the monitor.
If you have configured SHOCK 1, 2, and 3 to escalating energy levels, and then you manually
change the energy level outside preconfigured SHOCK 1, 2, 3 sequence and deliver a shock, it
disables the automatic energy escalation. See the Energy Level: Shock 1, 2, 3 section of the
R Series Configuration Guide for more details.
DEFIB 120J SEL.
Code
Marker
Options
00:01
ECG
Param
Report
Data
Sync
On/Off
Alarms