ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 54

R Series Service Manual
2-2
9650-0903-01 Rev. J
ALARM SET
ALARM SET status message when setting alarms.
ANALYSIS HALTED
ECG analysis halted due to user interaction such as:
• Lead/size change
• Analyze button was pressed again
• Impedance fault
• Charging error detected in auto defib mode
ASYNC PACE ONLY
Posted along with other faults. Indicates that the PD
module cannot detect sync pulses.
ATTACH PADS
AED: No pads on Auto Defib power-up.
Verify proper OneStep cable/hands-free therapy
electrode connection by disconnecting and
reconnecting the OneStep cable and hands-free
therapy-electrodes.
AUDIO QUEUE FULL
Indicates that the audio output queue is full. Additional
voice prompts can't be queued at this time.
None.
BATT HIGH CURRENT
Battery is not charged and battery current is greater than
1.6 A.
Replace battery or AC charger.
BATT HIGH VOLTAGE
Indicates that the charger voltage is too high.
Replace battery or AC charger.
BATT LOW VOLTAGE
Indicates that the charger voltage is too low.
Replace battery or AC charger.
BATTERY COMMS ERROR
Battery is not communicating with the host.
Replace battery interconnect board.
BATTERY FAULT
Replace battery.
BATTERY ID FAULT
Replace battery.
CAL. BARO. PRESSURE
Barometric pressure reading is out of range.
Calibrate the barometric pressure.
CALIBRATE NIBP
NIBP calibration is incomplete or failed.
Cycle power and retry; if problem persist calibrate
NIBP system.
CANNOT CHARGE
Cannot charge when charge button pressed.
Replace high voltage module or capacitor.
Error Message
Explanation
User
Advisory
Technical Action