Chapter 6 monitoring ecg – ZOLL Propaq MD Rev A User Manual
Page 79

9650-0805-01 Rev. A
Propaq MD Operator’s Guide
6-1
Chapter 6
Monitoring ECG
This chapter describes how to use the Propaq MD unit to monitor ECG.
Propaq MD units can perform ECG monitoring through 3-, 5-, or 12-lead ECG patient cables,
Multi-Function Pads, or standard defibrillation paddles. The use of an ECG patient cable and
electrodes is required, however, to monitor ECG during pacing.
You can use a 3-lead, 5-lead, or 12-Lead wire configuration for ECG monitoring (see Chapter
13, Monitoring 12-Lead ECG, for information on 12-lead monitoring).
Warning!
•
Excessive body hair or wet, sweaty skin may interfere with electrode adhesion.
Remove the hair and/or moisture from the area where the electrode is to be attached.
•
Use only electrodes that are well within the expiration date indicated on the package.
•
Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.
•
Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within
a few seconds.
•
To assure protection against the effects of defibrillator discharge, use only
ZOLL-approved accessories.