beautypg.com

Selecting default alarm limits – ZOLL Propaq MD Rev A User Manual

Page 76

background image

C

HAPTER

5 A

LARMS

5-8

www.zoll.com

9650-0805-01 Rev. A

Selecting Default Alarm Limits

The first three options --

Default Adult, Default Pediatric, Default Neonate

-- allow you to set all

alarm limits to the Propaq MD unit’s factory-specified default values, by patient type.

Warning!

A potential hazard exists if different alarm limits are used for the same or similar
equipment in any single area.

Confirm the alarm limits are appropriate for the patient each time there is a new
patient case.

Do not set alarm limits to such extreme values that render the alarm system useless.

Setting Alarm Limits Relative to the Patient -- Stat Set Option

The Propaq MD unit also allows you to set all alarm limits relative to the patient’s current vital
signs measurements by performing the following actions:

1. Press .

2.

Press the Alarm quick access key (

).

3.

Press the Stat Set quick access key (

). The Propaq MD unit sets all parameters to a new

value based on the current values as follows:

Parameter

(units)

Range

Upper Limit Calculation

Lower Limit Calculation

HR/PR

(bpm)

Numeric < 26

Limit is unchanged

Limit = 25

26

Numeric  99

Limit = Numeric x 1.2

Limit = Numeric x 0.8

100

 Numeric  250

Limit = Numeric + 20

Limit = Numeric – 20

Numeric > 250

Limit = 250

Limit is unchanged

IBP

(mmHg)

Numeric < 26

Limit = Numeric + 5

Limit = Numeric – 5

26

 Numeric  99

Limit = Numeric x 1.2

Limit = Numeric x 0.8

Numeric > 99

Limit = Numeric + 20

Limit = Numeric – 20

NIBP

(mmHg)

Numeric < 26

Limit = Numeric + 5

Limit = Numeric – 5

26

 Numeric  99

Limit = Numeric x 1.2

Limit = Numeric x 0.8

Numeric > 99

Limit = Numeric + 20

Limit = Numeric – 20

RR/BR

(/min)

Numeric < 26

Limit = Numeric + 5

Limit = Numeric – 5

26

 Numeric  99

Limit = Numeric x 1.2

Limit = Numeric x 0.8

Numeric > 99

Limit = Numeric + 20

Limit = Numeric – 20

Stat

Set