After transferring data, you must remove the USB drive and reinsert it before attempting
another transfer. If the USB device does not establish communication with the Propaq MD unit,
try powering off the unit and then on again to establish communication.

Clearing the Log

You should clear the patient data log after transferring data to the USB device or when the log is
full.

Note:

Clearing the log during patient treatment results in the loss of all patient data and
events recorded prior to clearing the log. Clearing the log creates a New Patient record
and all patient-specific parameters (alarm limit, defibrillation energy, etc.) are set to
their default values.

To clear the log:

1. Press .

2.

Press .

3.

Press the Clear Log quick access key (

).

3. Use the navigation keys to select Yes.

Warning!

To avoid a possible shock hazard, do NOT make any electrical connections to the USB
port except to connect a USB flash drive while in the vicinity of the patient.