Specifications

A-20

www.zoll.com

9650-0805-01 Rev. A

Respiration Rate Accuracy:

0 to 70 bpm:

±1 bpm

71 to 120 bpm: ±2 bpm

121 to 149 bpm: ±3 bpm

Flow rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.

Total System Response Time: 2.9 seconds typical, 3.9 seconds maximum.

Pulse Oximeter

Measurement Range: 0 to 100%

Accuracy: 70 to 100% +/- 3%, 0 to 70% unspecified

Note:

Since pulse oximeter equipment measurements are statistically distributed, only two-
thirds of pulse oximeter equipment measurements can be expected to fall within
+/- 3% of the value measured by a CO-oximeter.

Saturation Update Period: Updated once per second

SpO2 Saturation Alarm Hold Off: 10 seconds

Pulse Rate Range: 20 to 250 BPM ± 3% or 3 BPM whichever is greater

Pulse Rate Update Period: Updated once per second

Pulse Rate and Saturation Display:

Normal Response: Average of pulse rate and saturation signal over 6 to 7 seconds.

Wavelengths: Approximately 660nm and 900nm

Displayed Data:

Numeric blood oxygen saturation value

Numeric pulse rate

Pulse Amplitude bar indicator

Plethysmographic waveform

Probe Compatibility: DS-100A, D-25, R-15, D-25L, D-20, EC-4, EC-8, RS-10

Power Output: < 15 mW