Pulse oximeter – ZOLL Propaq MD Rev A User Manual
Page 224

Specifications
A-20
www.zoll.com
9650-0805-01 Rev. A
Respiration Rate Accuracy:
0 to 70 bpm:
±1 bpm
71 to 120 bpm: ±2 bpm
121 to 149 bpm: ±3 bpm
Flow rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.
Total System Response Time: 2.9 seconds typical, 3.9 seconds maximum.
Pulse Oximeter
Measurement Range: 0 to 100%
Accuracy: 70 to 100% +/- 3%, 0 to 70% unspecified
Note:
Since pulse oximeter equipment measurements are statistically distributed, only two-
thirds of pulse oximeter equipment measurements can be expected to fall within
+/- 3% of the value measured by a CO-oximeter.
Saturation Update Period: Updated once per second
SpO2 Saturation Alarm Hold Off: 10 seconds
Pulse Rate Range: 20 to 250 BPM ± 3% or 3 BPM whichever is greater
Pulse Rate Update Period: Updated once per second
Pulse Rate and Saturation Display:
Normal Response: Average of pulse rate and saturation signal over 6 to 7 seconds.
Wavelengths: Approximately 660nm and 900nm
Displayed Data:
Numeric blood oxygen saturation value
Numeric pulse rate
Pulse Amplitude bar indicator
Plethysmographic waveform
Probe Compatibility: DS-100A, D-25, R-15, D-25L, D-20, EC-4, EC-8, RS-10
Power Output: < 15 mW