Begin cpr following medical protocols, Prepare patient, Turn unit on – ZOLL Propaq MD Rev A User Manual
Page 168

M
ANUAL
D
EFIBRILLATION
14-6
www.zoll.com
9650-0805-01 Rev. A
Begin CPR Following Medical Protocols
Request additional assistance.
Prepare Patient
Remove all clothing covering the patient’s chest. Dry chest if necessary. If the patient has
excessive chest hair, clip or shave it to ensure proper adhesion of the electrodes.
Attach hands-free therapy electrodes according to instructions on the electrode packaging.
Ensure that the therapy electrodes are making good contact with the patient’s skin and are not
covering any part of the ECG electrodes.
Therapy Electrode Application
Warning!
Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.
1. Apply one edge of the pad securely to the patient.
2.
Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.
Note:
If it is not possible to place the “BACK” electrode on the patient’s back, place the
electrodes in the standard apex-sternum positions. Effective defibrillation results, but
pacing will usually be less effective.
Turn Unit On
Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.
If defibrillation electrodes are not making good contact with the patient’s skin and the pad
selection is ECG Lead, the unit issues the message CHECK THERAPY ELECTRODES and
does not allow the delivery of energy.
Pad
Skin