Invasive pressures, Temperature – ZOLL Propaq MD Rev A User Manual
Page 226
Specifications
A-22
www.zoll.com
9650-0805-01 Rev. A
Blood Pressure Validation:
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.
Invasive Pressures
Number of Channels: 3
Pressure range: -30 to 300 mmHg
Pressure Accuracy: +/- 2 mmHg or 2% of reading, whichever is greater, plus transducer error.
Pulse Rate Range: 25 to 250 BPM
Pulse Rate Accuracy: +/- 3 BPM, or +/- 3% of value whichever is greater
Pulse Rate Display: Average of last 4 beat-to-beat intervals.
Zero Adjust: + / - 200 mmHg
Transducer:
Sensitivity: 5uV/V/mmHg
Offset: +/- 125 mmHg including transducer offset
Excitation Impedance Range: 150 to 10,000 ohms
Excitation Voltage: 4.75 +/- 0.25 VDC
Connector: 6-pin circular MS3100 series
Temperature
Number of Channels: 2
Measurement Range: 0° to 50° C
Accuracy:
± 0.1° C from 10° C to 50° C
± 0.2° C from 0° C to 10° C
Resolution: 0.1° C
Scale: Fahrenheit or Celsius.
Temperature Display Signal: 20Hz, no averaging.
Probe: YSI 400 and 700 series
Display: T1, T2, T
Minimum Measurement Time: See the probe’s Instructions for Use to obtain minimum
measurement times for accurate readings. The Propaq MD does not add any clinically
significant time to obtain accurate readings.
Connect to:
A
B
C
D
E
Signal Type
Sig (-)
Exc (+)
Sig (+)
Exc (-)
shield