Using the spo, Parameter control panel, Setting upper and lower systolic alarm limits – ZOLL Propaq MD Rev A User Manual
Page 120

C
HAPTER
9 M
ONITORING
P
ULSE
O
XIMETRY
(S
P
O2)
9-8
www.zoll.com
9650-0805-01 Rev. A
6.
When you are finished changing values on the alarm menu, navigate to the Backarrow key
to exit the menu.
Setting Upper and Lower Systolic Alarm Limits
Initially, the SpO
2
Alarm Settings menu specifies whether the SpO
2
alarms are enabled (
ON
)
or
disabled (
OFF
),
and displays the default upper and lower systolic limits. The upper and lower
limits can be
ON
or
OFF
(default is
OFF
).The following table lists the default SpO
2
limits for
adult, pediatric, and neonate patients, and gives the range in which you can set these limits.
Using the SpO
2
Parameter Control Panel
Use the navigation keys to highlight and select the SpO
2
numeric display to display the
SpO
2
Parameter Control Panel, on which you can select options that optimize the use of SpO
2
for the patient:
Figure 9-4 SpO
2
Parameter Control Menu
Patient Type
SpO
2
Limit Default
SpO
2
Limit Range
Adult
Lower: 85%
Upper: 100%
Lower: 50 - 98%
Upper: 52 - 100%
Pediatric
Lower: 85%
Upper: 100%
Lower: 50 - 98%
Upper: 52 - 100%
Neonate
Lower: 85%
Upper: 95%
Lower: 50 - 98%
Upper: 52 - 100%