Printing 12-lead waveform traces – ZOLL Propaq MD Rev A User Manual
Page 159
Printing 12-Lead Waveform Traces
9650-0805-01 Rev. A
Propaq MD Operator’s Guide
13-9
Printing 12-Lead Waveform Traces
Once you have determined that the Propaq MD unit is correctly setup for 12-Lead ECG
monitoring, you can print the 12-Lead ECG traces for review and analysis.
Press the Acquire
quick access key (
) to collect 10 seconds of 12-Lead data for print.
Press the Print quick access key (
) to print a 12-lead snapshot, which begins with a header
that gives the date, time, and patient information, followed by 3.3 second samples of all twelve
waveform traces. The waveform traces print in either a 3 x 4 format (the default).
or a 2 x 6 format:
The Propaq MD unit stores a minimum of 32 12-Lead snapshots in a separate log. Once 32 12-
lead snapshots are stored, the oldest snapshot in the log is overwritten by subsequent snapshots.
Note that if a patient data log is full, you will not be able to store12-lead snapshots. The 12-lead
snapshots are erased when you clear the log.
When you are done viewing and printing the 12-Lead waveform traces, press the Exit 12 quick
access key (
)to restore the display of other monitoring functions.
12
12
Exit