12 While the defibrillator is still charged, press

the paddles electrodes together then press
and hold the SHOCK button on the Sternum
paddle (only).

Verify that the defibrillator does not
discharge.

13 While the defibrillator is charged, press the

paddles electrodes together then
simultaneously press and hold both the
Apex and Sternum paddle SHOCK buttons.

Verify that the defibrillator discharges and
that a DEFIB SHORT TEST PASSED
message displays.

If the message DEFIB SHORT TEST FAILED
displays, contact the appropriate technical
personnel or the ZOLL Technical Service
Department.