Pacer

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

A-19

Pacer

Type: External transcutaneous pacing

Pacer Rate: 30 to 180 BPM ± 1.5%.

Output Current: 0 to 140 mA ± 5% or 5 mA (whichever is greater)

Modes: Demand and Fixed

Status Indicators:

ECG lead fault, pace marker on monitor and chart, start/stop indicator on
display.

Pulse Type: Rectilinear, constant current

Pulse Width: 40 ms +/-2 ms

CO

2

Range: 0 to 150 mmHg

Accuracy CO

2

:

* At sea level.

** Accuracy applies for breath rates of up to 80 bpm. For breath rates above 80 bpm,
accuracy is 4 mmHg or ±12% of reading whichever is greater, for EtCO2 values exceeding
18 mmHg. This is tested according to and is compliant with ISO 21647. To achieve the
specified accuracies for breath rates above 60 breaths/minute, the Microstream FilterLine
H Set for Infant/Neonatal must be used. Above 40 C, ± 1mmHg or ± 2.5% (whichever is
greater) has to be added to the tolerance of the accuracy specs.

Drift of Measurement Accuracy: Over any 24-hour period, the accuracy claims listed above
are maintained.

The accuracy specification is maintained to within 4% for the following gas mixtures (all
values are in Vol.%).

Respiration Range: 0 to 149 breaths per minute

CO2 Partial Pressure*

Accuracy **

0-38 mmHg

± 2 mmHg

39-99 mmHg

± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)

100-150 mmHg

± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)

CO

2

N

2

O

2

N

2

O

H

2

0

Anesthetic Agents

1.0 to 13

0 to 97.5

0 to 100

0 to 80

Dry to saturated

According to EN 21647