ZOLL Propaq MD Rev A User Manual
Page 107
Configuring NIBP Alarms and Settings
9650-0805-01- Rev. A
Propaq MD Operator’s Guide
8-11
5.
On the selected NIBP alarm settings menu, use the navigation keys to select the fields that
you want to change. The fields are
•
Upper Limit Enable
•
Lower Limit Enable
•
Upper Limit
•
Lower Limit
6.
When you are finished changing values on the alarm settings menu, select the back arrow to
exit the menu.
Setting Upper and Lower Systolic Alarm Limits
Initially, the NIBP Systolic Alarm Settings menu specifies that the NIBP systolic pressure
alarms are enabled (
ON
)
or disabled (
OFF
),
and displays the default upper and lower systolic
limits. The upper and lower limits can be
ON
or
OFF
(default is
OFF
). The following table lists
the default NIBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Setting Upper and Lower Diastolic Alarm Limits
Initially, the NIBP Diastolic Alarm Settings menu specifies that the NIPBP diastolic pressure
alarms are enabled (
ON
)
or disabled (
OFF
),
and displays the default Upper and Lower diastolic
limits. The upper and lower limits can be
ON
or
OFF
(default is
OFF
). The following table lists
the default diastolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Patient Type
NIBP Systolic Limit Default
NIBP Systolic Limit Range
Adult
Lower: 75 mmHg
Upper: 220 mmHg
Lower: 30-258 mmHg
Upper: 32-260 mmHg
Pediatric
Lower: 75 mmHg
Upper: 145 mmHg
Lower: 30-158 mmHg
Upper: 32-160 mmHg
Neonate
Lower: 50 mmHg
Upper: 100 mmHg
Lower: 20-118 mmHg
Upper: 22-120 mmHg
Patient Type
Diastolic Limit Default
Diastolic Limit Range
Adult
Lower: 35 mmHg
Upper: 110 mmHg
Lower: 20-218 mmHg
Upper: 22-220 mmHg
Pediatric
Lower: 35 mmHg
Upper: 100 mmHg
Lower: 20-128 mmHg
Upper: 22-130 mmHg
Neonate
Lower: 30 mmHg
Upper: 70 mmHg
Lower: 10-108 mmHg
Upper: 12-110 mmHg