Displaying SpO

2

Measurements

9650-0805-01 Rev. A

Propaq MD Operator’s Guide

9-7

3.

Lower the clear plastic cover over the connection to secure it:

Figure 9-3 Lower Plastic Cover Over Cable Connection

Displaying SpO

2

Measurements

When the connection is made between the SpO

2

sensor and the unit, the message,

INITIALIZING, appears in the SpO

2

display. After a brief delay, the unit displays the SpO

2

measurement. If the message, SENSOR FAILURE, appears, the sensor is either incompatible
with the Propaq MD unit, or it is not working, and you will need to replace the sensor.

A blip bar appears on the right side of the SpO

2

numeric display window. The blip bar indicates

the relative pulsatile strength and quality of the SpO

2

data.

Enabling/Disabling SpO

2

Alarms and Setting Alarm Limits

When enabled, the Propaq MD unit sounds alarms whenever measurements are outside set
limits for the High and Low SpO

2

values.

You can enable (or disable) SpO

2

alarms and set Upper and Lower alarm limits through the

Alarms quick access key (

)or through the SpO

2

Parameter Control Panel.

To configure SpO

2

alarms through the Alarms quick access key:

1. Press the More quick access key (

).

2.

Press .

3.

Press the Limits quick access key (

).

4.

Use the navigation buttons to highlight and select the appropriate alarm menu selection. For
NIBP, the alarm menu selection is SpO

2

Alarm.

5.

On the selected SpO

2

Alarm Settings menu, use the navigation keys to select the fields that

you want to change. The fields are

Upper Limit Enable

Lower Limit Enable

Upper Limit

Lower Limit

Limits