E

XTERNAL

P

ACING

15-6

www.zoll.com

9650-0805-01 Rev. A

Therapy Electrode Application

Warning!

Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.

1. Apply one edge of the pad securely to the patient.

2.

Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.

3.

Ensure that hands-free therapy electrodes are making good contact with the patient’s skin and
are not covering any part of any other ECG electrodes.

4.

Press the Lead quick access key and select I, II, or III to provide the largest amplitude QRS
complex.

Note:

When the Pacer is on, the lead selection is restricted to Leads I, II, or III.

5.

Verify that R-waves are being properly detected by confirming that a QRS tone occurs with
each displayed R wave or by verifying that the Propaq MD unit’s heart rate display accurately
reflects the patient’s pulse rate.

3 Press Pacer button

Press the Pacer button on the front panel of the unit. The Pacer Settings window displays.

4 Set Mode

Use the arrow keys to navigate to Mode, press the Select button, then use the arrow keys and
Select button to set the Pacer Mode to Fixed. The selected mode is displayed in the pacing
window.

Note:

The initial pacing mode can be set in Defib/Pacer Default settings in the Supervisor
Setup menu.

Pad

Skin

Pacer Settings

Mode

Rate

Output

Start Pacer

Turn Pacer Off

Demand

60