Determine the Patient’s Condition Following Local Medical Protocols

Verify the following:

Unconsciousness

Absence of breathing

Absence of pulse

Begin CPR Following Local Medical Protocols

Request additional assistance.

Turn Unit On

Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.

1 Select Energy Level

Press the Select Energy arrows up or down to select the desired energy level. These buttons are
located either on the front of the unit or on the STERNUM paddle.

Note:

The initial defibrillator pad/external paddle energy selection can be set in Defib/Pacer
default settings in the Setup>Supervisor menu.

The following are the default energy selections:

Adult - 120 joules

Pediatric- 50 joules

Neonate - 50 joules

Note:

Neonatal and pediatric defibrillator energy levels should be selected based on site-
specific protocols.

The selected energy level is displayed at the bottom of the display screen.

Prepare Paddles

Ensure that the paddles are connected to the multi-function cable (MFC) and that the MFC is
connected to the Propaq MD unit. Apply a liberal amount of electrolyte gel to the electrode
surface of each paddle, and rub the electrode surfaces together to evenly distribute the applied
gel. (You can substitute electrode gel patches for the gel.)