Nibp system messages – ZOLL Propaq MD Rev A User Manual
Page 111

NIBP System Messages
9650-0805-01- Rev. A
Propaq MD Operator’s Guide
8-15
NIBP System Messages
When monitoring NIBP, the Propaq MD unit may display the following messages:
System Message
Cause
READING IN PROGRESS
The unit is taking an NIBP measurement and
functioning normally.
READING STOPPED
The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP button
and cancelled the measurement.
READING FAILED
The patient’s pulse is too weak to for NIBP
measurement, or the cuff requires adjustment.
ARTIFACT
Excessive artifact is preventing the NIBP
measurement.
NEONATE ALERT
The unit has detected a neonate cuff when in Adult
mode. Replace cuff or correct patient mode, as
appropriate.
AIR LEAK
A major air leak is preventing cuff inflation. Check
hose and cuff connections, replace a defective hose
or cuff, as necessary, and reattempt NIBP
measurement.
KINKED HOSE
An air obstruction is preventing the correct operation
of the unit. Check hose for kinks. Eliminate kinks or
replace hose, if necessary.
NIBP DISABLED
A system error has occurred, and the Propaq MD
unit should be serviced.
NIBP CALIBRATION CHECK DUE The annual calibration check is due. You can
continue to use the device, however, we recommend
that you contact the ZOLL Technical Service
Department or a ZOLL authorized service
representative as soon as possible to perform the
recommended annual NIBP calibration check.