Specifications
A-18
www.zoll.com
9650-0805-01 Rev. A
General
Weight:
9.9 lbs.without battery and paper
11.6 lbs.with battery and paper
Dimensions:
Without Handle: 8.9” x 8.75” x 7.9”
With Handle: 8.9” x 10.4” x 7.9”
Operating:
Temperature: 0 to 50° C
60° C for 6 hours after storage at room temperature
Humidity: 15 to 95% RH (non-condensing)
Vibration:
•
MIL-STD 810G, Method 514.6, Profile for combined UH-1, UH-60, and CH-47
Rotary Wing Aircraft.
•
EN 1789 for ambulance.
Shock: MIL-STD 810G, Method 516.6, Tested at 75 g
Drop: MIL-STD 810G, Method 516-6, Tested at 1 meter with 26 drops
IEC 60601-1, Tested at 2 meters
Altitude: 170 M to 4572 M (-557 feet to 15,000 feet)
Transport and Storage:
Temperature: -30 to 70°C
Note:
The Propaq MD device may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.
Humidity: 15 to 95% RH (non-condensing)
Atmospheric pressure: 572 mbar to 1034 mbar
Shock/vibration: ISTA 1A
Safety Classification: Class 1 and internal power per IEC/EN 60601-1
Enclosure Protection:
Solid Foreign Object: IEC 60529, IP5X
Water: IEC 60529, IPX5
Auxiliary Operating Power:
Propaq MD Auxiliary Power Adapter 8300-0004
Input:
100-240V
50-60 Hz, 2A
100-115V
400 Hz, 2A
Output: 14.5V
4.15A
80W (peak)
IP Rating: IP23