Configuring the Waveform Display
9650-000820-01 Rev. B
Propaq M Operator’s Guide
3-7
Configuring the Waveform Display
You can display up to four waveform traces on the Waveform Display window. The first
waveform trace always uses an ECG lead as its source (such as Leads
l
,
ll
, or
lll
, and so on). As
you insert the remaining three traces, you can specify that the traces use an ECG lead as the
waveform source, or that the trace derive its waveform from other available monitoring
functions (such as
Resp
,
CO2
,
SpO2
or IBP channels
P1
,
P2
, or
P3
).
The Propaq M unit can also cascade a trace onto the adjoining trace area to double the duration
of the trace display.
On the Waveform Display window, to insert a new trace (
Insert
) or cascade (
Cascade
) a
displayed trace, highlight and select the trace label above the trace. In the following example,
the unit is configured to cascade the ECG Lead l trace:
Note:
The Propaq M unit automatically inserts a new waveform when you turn on a
parameter (CO
2
) or a new sensor signal is present (SPO
2
, IBP). The Propaq M unit
automatically removes a waveform when you turn off a parameter or remove a sensor
and the unit displays the resulting equipment alert.