beautypg.com

Ready for use (rfu) indicator – ZOLL Propaq M Rev B User Manual

Page 19

background image

Propaq M Product Functions

9650-000820-01 Rev. B

Propaq M Operator’s Guide

1-9

After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.

Ready For Use (RFU) Indicator

The Propaq M has an RFU indicator on the front panel that indicates if the device is ready for
use. The RFU indicator has three states which are described in the following table.

State

Description

Action

Ready for Use

The device is ready for use. Patient
monitoring is functional and the
battery is above the low battery
capacity.

Note: If the device is plugged into
the auxiliary power adapter, the
Ready for Use indicator may display
even if the battery is depleted.
Check the status of the battery
before removing the device from the
auxiliary power adapter.

None required.

Flashing

One or more of the following has
occurred:

The battery is not properly
installed.

A low battery is installed.

A battery fault has occurred.

There is no battery installed
while connected to auxiliary
power.

One or more patient monitoring
parameters have failed self-test
(NIBP, SpO

2

, CO

2

, IBP, or

Temp).

The front panel button self-test
failed.

The speech database self-test
failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to flash, remove the unit
from service and contact the
appropriate technical personnel or
the ZOLL Technical Service
Department.

Do Not Use

One or more of the following has
occurred:

The battery is not properly
installed.

No battery is installed and
auxiliary power is not present.

A very low battery (below
software shutdown limit) was
installed.

ECG self-tests have failed, or
other critical self-tests have
failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to display the Do Not
Use symbol, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.