Specifications

A-6

www.zoll.com

9650-000820-01 Rev. B

Enclosure Protection:

Solid Foreign Object: IEC 60529, IP5X

Water: IEC 60529, IPX5

Auxiliary Operating Power:

Auxiliary Power Adapter, 8300-0004

Input:

100-240V

50-60 Hz, 2A

100-115V

400 Hz, 2A

Output: 14.5V

4.15A

80W (peak)

IP Rating: IP23

CO

2

Range: 0 to 150 mmHg

Accuracy CO

2

:

* At sea level.

** Accuracy applies for breath rates of up to 80 bpm. For breath rates above 80 bpm,
accuracy is 4 mmHg or ±12% of reading whichever is greater, for EtCO

2

values exceeding

18 mmHg. This is tested according to and is compliant with ISO 21647. To achieve the
specified accuracies for breath rates above 60 breaths/minute, the Microstream FilterLine
H Set for Infant/Neonatal must be used. Above 40 C, ± 1mmHg or ± 2.5% (whichever is
greater) has to be added to the tolerance of the accuracy specs.

Drift of Measurement Accuracy: Over any 24-hour period, the accuracy claims listed above
are maintained.

The accuracy specification is maintained to within 4% for the following gas mixtures (all
values are in Vol.%).

Respiration Range: 0 to 149 breaths per minute

Respiration Rate Accuracy:

0 to 70 bpm:

±1 bpm

71 to 120 bpm: ±2 bpm

121 to 149 bpm: ±3 bpm

Flow rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.

Total System Response Time:

2.9 seconds typical, 3.9 seconds maximum.

CO

2

Partial Pressure*

Accuracy **

0-38 mmHg

± 2 mmHg

39-99 mmHg

± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)

100-150 mmHg

± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)

CO

2

N

2

O

2

N

2

O

H

2

0

Anesthetic Agents

1.0 to 13

0 to 97.5

0 to 100

0 to 80

Dry to saturated

According to EN 21647