Cautions
9650-000820-01 Rev. B
Propaq M Operator’s Guide
1-15
Cautions
If the unit is to be stored longer than 90 days, remove the battery pack.
Do not sterilize the monitor, or its accessories unless the accessories are labelled as sterilizable.
Do not immerse any part of the monitor in water.
Do not use the monitor if excessive condensation is visible on the device. Wipe only the outside
with a damp cloth.
Do not use ketones (such as acetone or MEK) on the monitor.
Avoid using abrasives (including paper towels) on the display window.
To achieve the specified level of protection against spilled or splashed liquids, thoroughly dry
all exposed surfaces of this device prior to operation or connections to auxiliary power.
If liquids enter the device connectors, remove all liquid from the connectors and allow the
device to dry thoroughly prior to use.
Grounding reliability can be achieved only when the equipment is connected to a receptacle
marked “HOSPITAL ONLY,” “HOSPITAL GRADE,” or equivalent. If the grounding integrity
of the line cord or ac receptacle is questionable, operate the monitor using battery power only.
Do not connect to an electrical outlet controlled by a wall switch or dimmer.
For accurate ECG information, and to protect against noise and other interference, use only
internal current-limiting ECG cables specified or supplied by ZOLL.
For continued safety and EMI performance, use only the line cord supplied by ZOLL.
Electrical installation of the room or the building in which the monitor is to be used must
comply with regulations specified by the country in which the equipment is to be used.
Dispose of battery packs in accordance with national, regional and local regulations. Battery
packs should be shipped to a reclamation facility for recovery of metal and plastic compounds
as the proper method of waste management.