beautypg.com

Operator safety, Patient safety – ZOLL Propaq M Rev B User Manual

Page 24

background image

C

HAPTER

1 G

ENERAL

I

NFORMATION

1-14

www.zoll.com

9650-000820-01 Rev. B

Operator Safety

Do not use the Propaq M in the presence of oxygen-rich atmospheres, flammable anesthetics,
or other flammable agents (such as gasoline). Using the unit in such environments might cause
an explosion.

Do not use the unit near or within standing water. Electrical safety might be compromised when
the monitor is wet.

The use of accessory equipment that does not comply with the equivalent safety requirements
of the Propaq M monitor could reduce the level of safety of the combined system. When
choosing accessory equipment, consider the following:

Use of the accessory in the patient vicinity.

Evidence that the safety certification of the accessory has been performed in accordance
with the appropriate IEC (EN) 60601-1 and/or IEC (EN) 60601-1-1 harmonized national
standards.

Always check that the equipment functions properly and is in proper condition before use.

Patient Safety

This equipment should be connected to only one patient at a time.

To ensure patient safety, do not place the monitor in any position that might cause it to fall on
the patient.

To ensure patient safety, connect the Propaq M only to equipment with circuits that are
electrically isolated.

Use only high-quality ECG electrodes.

Do not use ECG electrodes if the gel is dried, separated, torn or split from the foil; patient burns
may result from using such electrodes.

Check the expiration date on the electrode packaging. Do not use electrodes after their
expiration date.

Excessive body hair or wet, diaphoretic skin can inhibit electrode coupling to the skin. Clip
excess hair and dry any moisture from the area where an electrode is to be attached.

Carefully route the patient cables away from the patient’s neck to reduce the possibility of
patient entanglement or strangulation.

To avoid electrosurgery burns at monitoring sites, ensure proper connection of the
electrosurgery return circuit so that a return path cannot be made through monitoring electrodes
or probes.

During electrosurgery, observe the following guidelines to minimize electrosurgery unit (ESU)
interference and provide maximum operator and patient safety:

Keep all patient monitoring cables away from earth ground, ESU knives, and ESU return
wires.

Use electrosurgical grounding pads with the largest practical contact area.

Always ensure proper application of the electrosurgical return electrode to the patient.

Check electrical leakage levels before use. Leakage current might be excessive if more than one
monitor or other piece of equipment is connected to the patient.