9650-000820-01 Rev. B

Propaq M Operator’s Guide

13-1

Chapter 13

12-Lead ECG Interpretive Analysis

The Propaq M 12-Lead input is Type CF defibrillator proof.

This chapter describes how to use the Propaq M unit to monitor 12-Lead ECG for adult and
pediatric patients, and how to display 12-Lead ECG Interpretive Analysis information for adult
patients.

Propaq M 12-Lead ECG Monitoring provides simultaneous acquisition and storage of
12-Lead information for adult and pediatric patients, and post-acquisition Interpretive Analysis
for adult patients.

Warning!

12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals from
adult and pediatric patients in the supine, resting position -- always ensure that the
patient is kept motionless during 12-lead ECG signal acquisition and analysis. Use of
the device to acquire ECG signals from moving or shaking patients may produce noisy
signals that are difficult to interpret.

12-Lead Interpretive Analysis is for use with only adult patients.

Excessive body hair or wet, sweaty skin may interfere with electrode adhesion. Remove
the hair and/or moisture from the area where the electrode is to be attached.

Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.

Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within a
few seconds.