Cleaning the Propaq M unit
To clean the Propaq M unit, use a nearly dry cloth containing one of the mild cleaning agents
listed below. DO NOT allow cleaning agent or water to run into the crevices or connector
openings at any time. Thoroughly wipe off any excess cleaning solution from the Propaq M
unit with a dry cloth. Always check monitor and connector opening for unusual wear, damage
or dampness while cleaning
Use only these recommended cleaning agents:
•
Warm water
•
Hydrogen peroxide solution
•
Liquid soap
•
Wex-cide
®
•
Windex
®
•
T.B.Q.
®
Never use these cleaning agents:
•
Butyl alcohol
•
Denatured ethanol
•
Freon
•
Mild chlorine bleach solution
•
Isopropyl alcohol
•
Trichloroethane, trichloroethylene
•
Acetone
•
Vesphene II
•
Enviroquat
•
Staphene
•
Misty
•
Glutaraldehyde
Cleaning the NIBP Blood Pressure Cuff
Clean the cuff with common hospital disinfectants, including Clorox
®
(1:10 solution),
isopropyl alcohol, Lysol
®
solution, Phisohex
®
, Quatricide
®
, Virex
®
and Vesphene
®
. Wash
gently with the solution, then rinse. DO NOT allow solution to enter cuff tubes. Allow the cuff
and hose to completely dry before patient use.
Cleaning SpO
2
Sensors
Clean the SpO
2
sensors with a cloth that has been slightly dampened with one of the
recommended agents listed above. DO NOT submerge the probe or its connector in any liquids
or cleaning agents. Thoroughly wipe off any excess cleaning solution with a dry cloth. Allow
the sensor to completely dry before patient use.