Cleaning cables and accessories – ZOLL Propaq M Rev B User Manual
Page 189
![background image](/manuals/355913/189/background.png)
Recommended Minimum Preventive Maintenance Schedule
9650-000820-01 Rev. B
Propaq M Operator’s Guide
16-3
Cleaning Cables and Accessories
Cables, cuff tubing, and other accessories can be wiped clean with a damp cloth moistened in a
mild detergent solution or according to manufacturer's instructions.
Recommended Minimum Preventive Maintenance
Schedule
Operational Tests should be performed at regular intervals. The Operational Tests augment the
automated self-tests that the Propaq M unit performs to help ensure readiness. The following
Operational Tests may be performed by the user:
•
Keypad Test
•
LED Test
•
LCD Test
•
RFU Indicator Test
•
Audio Test
Additional Tests may be performed by a qualified Biomedical Technician (BMET):
•
Battery Test
•
Fan Test
•
NIBP Test
•
USB Test
The Operational Test Procedures are described in the Propaq M Service Manual.
In addition to inspection and cleaning instructions, the following is recommended:
•
Perform NIBP calibration check
•
Perform CO
2
calibration check
•
Perform Functional Verification Tests
These tests shall be performed every year or according to local requirements (to be performed
by a qualified Biomedical Equipment Technician (BMET)) as described in the Propaq M
Service Manual.