Battery
9650-000820-01 Rev. B
Propaq M Operator’s Guide
A-5
Battery
Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh
Capacity:
With a new, fully charged battery operating at room temperature:
•
At least
7.5 hours of continuous monitoring of ECG, SpO
2
, CO
2
, three Invasive Pressure
channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes
(display set to 30%).
Note:
Proper battery care is required to maintain maximum available capacity.
Battery Indicators:
5 Battery capacity LED indicators, Fault indicator, Recalibration indicator
Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.
General
Weight:
Without Battery: 7.2 lbs.
With Battery: 8.5 lbs.
Dimensions:
Without Handle: 9.0” x 8.7” x 6.5”
With Handle: 9.0” x 10.4” x 7.1”
Operating:
Temperature: 0 to 50° C
Humidity: 15 to 95% RH (non-condensing)
Vibration:
•
MIL-STD 810G, Method 514.6, Profile for combined UH-1, UH-60, and CH-47
Rotary Wing Aircraft.
•
EN 1789 for ambulance
Shock: MIL-STD 810G, Method 516.6, Tested at 75 g
Bump: EN 1789 (IEC 60068-2-29)
Drop: MIL-STD 810G, Method 516-6, Tested at 1 meter with 26 drops
IEC 60601-1, Tested at 2 meters
Altitude: -170 M to 4572 M (-557 feet to 15,000 feet)
Transport and Storage:
Temperature: -30 to 70°C
Humidity: 15 to 95% RH (non-condensing)
Atmospheric pressure: 572 mbar to 1034 mbar
Shock/vibration: MIL STD 810G, Method 514.6, 4.4.2, Procedure II
Safety Classification: Class 1 and internal power per IEC/EN 60601-1
Note:
The Propaq M device may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.