WaveAce Remote Control
2. From the WaveAce menu, choose Utility > IP Settings (on the third page of settings).
3. Enter an IP Address, Subnet Mask, and Gate Way. Enter values by selecting each segment and
turning the Adjust knob until the desired number is reached. Press the top soft key to jump to the
next field or press the Adjust knob to tab to the next segment of the current field. Leave DHCP
disabled.
Program Messages
Program messages are composed of commands or queries separated by
semicolons
and ending with a
terminator
:
Command/Query Syntax
The general form of a command or a query is an optional
header path
, followed by a command
header
,
optionally followed by one or several
parameters
(shown as
in the following construct):
[header_path:]
NOTE:
There is a space between the header and the first parameter.
Commas separate parameters.
The question mark [?] is optional and turns the command into a query.
The oscilloscope does not distinguish between upper- and lowercase characters. Commands/queries are shown
mixed-case in this manual to call out the short form within the long form.
6