beautypg.com

Acquisition - trig_select, trse, Command syntax, Query syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 33: Response format, Example

background image

Operator's Manual

ACQUISITION - TRIG_SELECT, TRSE

The TRIG_SELECT command selects the condition that triggers the acquisition of waveforms. Only the

Edge trigger is supported.

Command Syntax

TRig_Select EDGE,SR,,HT,TI,HV,

:= {C1,C2,C3,C4,LINE,EX,EX5}

:=

Refer to the trigger information provided for your oscilloscope model.

NOTE: When specifying , the unit S (seconds) is optional.

Query Syntax

TRig_SElect?

Response Format

TRig_SElect ,SR,,HT,,HV,

HV2 only returned if is P2 or I2.

Example

The following instruction selects the single-source trigger with Channel 1 as trigger source. Hold type

and hold value are chosen as "pulse smaller" than 20 ns:

TRSE SNG,SR,C1,HT,PS,HV,20 NS

920836 Rev B