WaveAce Remote Control

STATUS - ALL_STATUS?, ALST?

The ALL_STATUS? query reads and clears the contents of all status registers: ESR, INR, DDR, CMR, and

EXR. For an interpretation of the contents of each register, refer to the appropriate status register.
The query is useful to obtain a complete overview of the state of your oscilloscope.

Query Syntax

ALl_STatus?

Response Format

ALl_STatus ESR,,INR,,DDR,,CMR,,
EXR,,URR,

:= 0 to 65535

Example

The following instruction reads the contents of all the status registers:

ALST?

Response message:

ALST ESR,000052,INR,000005,DDR,000000,CMR,000004,EXR,000024

Related Commands

*CLS, CMR?, DDR?, *ESR?, EXR?