Cursor - cursor_value?, crva, Query syntax, Response format – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 39: Example

background image

Operator's Manual

CURSOR - CURSOR_VALUE?, CRVA?

The CURSOR_VALUE? query returns the values measured by the specified cursors. The

PARAMETER_VALUE? query is used to obtain measured waveform parameter values.
The keyword ALL should not be used; neither should multiple keywords. If they are used, the word

UNDEF is returned.
For the CRVA? query to work, the specified trace must be visible, and the current cursor mode must be

the same as in the query. If it is not the same, UNDEF is returned.

Query Syntax

:CuRsor_VAlue? []

:= {TA,TB,TC,TD,C1,C2,C3,C4}

:= {HABS, HREL, VABS, VREL}

Where:

HABS

- Horizontal absolute

HREL

- Horizontal relative

VABS

– Vertical absolute

VREL

– Vertical relative

Response Format

:CuRsor_VAlue HABS,,

:CuRsor_VAlue HREL,,,,
,

The dV/dt value is displayed in the appropriate trace label box.

:CuRsor_VAlue VABS,

:CuRsor_VAlue VREL,

For horizontal cursors, both horizontal and vertical values are given. For vertical cursors only vertical

values are given.

Example

The following query reads the measured absolute horizontal value of the cross-hair cursor (HABS) on

Channel 2:

C2:CRVA? HABS

Response message:

C2:CRVA HABS,34.2E-6 S, 244 E-3 V

920836 Rev B