STATUS - EXR?
The EXR? query reads and clears the contents of the EXecution error Register (EXR). The EXR register
specifies the type of the last error detected during execution.
Query Syntax
EXR?
Response Format
EXR?
Example
The following instruction reads the contents of the EXR register:
EXR?
Response message (if no fault):
EXR 0
Related Commands
ALL_STATUS, *CLS
EXR? Status Register Structure
Value Description
21
Permission error. The command cannot be executed in local mode.
22
Environment error. The oscilloscope is not configured to correctly process a command. For instance,
the oscilloscope cannot be set to RIS at a slow timebase.
23
Option error. The command applies to an option which has not been installed.
24
Unresolved parsing error.
25
Parameter error. Too many parameters specified.
26
Non-implemented command.
27
Parameter missing. A parameter was expected by the command.
30
Hex data error. A non-hexadecimal character has been detected in a hex data block.
31
Waveform error. The amount of data received does not correspond to descriptor indicators.
32
Waveform descriptor error. An invalid waveform descriptor has been detected.
33
Waveform text error. A corrupted waveform user text has been detected.
34
Waveform time error. Invalid RIS or TRIG time data has been detected.