WaveAce Remote Control

<164> MAX_VALUE: float ; maximum allowed value. It corresponds

; to the upper edge of the grid.

;

<168> MIN_VALUE: float ; minimum allowed value. It corresponds

; to the lower edge of the grid.

;

<172> NOMINAL_BITS: word ; a measure of the intrinsic precision

; of the observation: ADC data is 8 bit

; averaged data is 10-12 bit, etc.

;

<174> NOM_SUBARRAY_COUNT: word ; for Sequence, nominal segment count

; else 1

;

<176> HORIZ_INTERVAL: float ; sampling interval for time domain

; waveforms

;

<180> HORIZ_OFFSET: double ; trigger offset for the first sweep of

; the trigger, seconds between the

; trigger and the first data point

;

<188> PIXEL_OFFSET: double ; needed to know how to display the

; waveform

;

<196> VERTUNIT: unit_definition ; units of the vertical axis

;

<244> HORUNIT: unit_definition ; units of the horizontal axis

74