Quick Start Reference – Calibrating Density

Step 1

Start at the Status Display

From the Status Display Screen, Press the F3 Key to display the
Password Screen (or the Main Menus if the password is disabled).

C

F1

F2

F3

F4

Ronan Engineering
X96S

Density Gage

Enter Password:

*

*****

All0

RST0

EXIT

ENTER

C

F1

F2

F3

F4

Ronan Engineering

X96S Density Gage
Status Display

D1

12.1860 % sol-wt

D2

1.1860 SGU

Proc Temp 50.00 deg
Raw Counts 12000

Filt.Counts 12000

Time 04/20/05 08:05
Diagnostics

BACK

Step 2 Password Menu

With the left most character * highlighted (All digits are set at
000000 at this point), enter in the password ‘101010’.
Press the

key 1 time to make the left most character equal to 1.

Press the

key 2 times to move to the third character.

Press the

key 1 time to make the third character equal to 1.

Press the

key 2 times to move to the fifth character.

Press the

key 1 time to make the fifth character equal to 1.

Press the F4 key (enter) to accept the password.
This will take you to the Main Menu.

Note: If the wrong password was entered, press the F1 (All0) to set
all the characters back to a value of 000000. You can begin to re-
enter the password from the beginning. Pressing the F2 (RST0) will
set the individual character that is highlighted back to the value 0.

Note: For security reasons, each character will always be displayed
as an asterisk

*

.