40

Quick Start Reference

Calibrating Density

What type of calibration is needed?

Two types of calibration are available for Ronan’s density gage.

1

.

Single-point calibration

This requires an accurate laboratory analysis of a truly representative sample of one process density. Single-Point
Calibration is used as an alternative method when it is not physically or economically practical to vary the process
density. In these cases, a second point is estimated based on the mathematical formula for radiation
transmission/absorption.

2. Dual-point calibration

This requires an accurate laboratory analysis of truly representative samples of two process densities. Dual-Point
Calibration is preferred when the process can be varied to obtain two process densities (one at each end of the
measurement range).

Important facts to remember –

• Equipment should be running at normal environmental conditions – temperature, pressure, process

flow, etc.

• Final gage measurements can only be as accurate as your sampling and analysis technique.

• Remember to document all changes made to the factory parameters that are active in your system.