Pressure reference menu, Auto reference menu, Diagnostics – Ronan X96S DENSITY GAGE User Manual
Page 33
Density Gage Manual 012407
29
Pressure Reference Menu
This menu is used to perform the process pressure reference procedure. You will be asked to provide real or simulated pressure
input conditions, and then enters the pressures relating to these conditions. It contains the following items:
ITEM
FUNCTION
Pressure Ref
This item invokes a method that performs the two point process pressure reference
procedure
Pres Cap Low
Shows the raw captured pressure counts for the first calibration point
Pres Cap High
Shows the raw captured pressure counts for the second calibration point
Auto Reference Menu
This menu is used to setup the auto reference mechanism. It contains the following items:
ITEM
FUNCTION
Auto Ref
Shows and allows the user to set the enable/disable flag that controls the auto reference
mechanism
Ref Delay
Shows and allows the user to set the delay between auto-reference signal and starting
the reference measurement
Auto Ref Density
Shows and allows the user to set the density for the auto-reference function
Diagnostics
ITEM
FUNCTION
Raw Counts
Non-filtered counts from the detector
Filt Counts
X96S filtered counts from the detector
Raw Dens.
Displays the real time raw density value
Current
Displays the real time 4-20ma value
Ref/Cal Data
Displays the information on the reference/calibration data
Alarms
Information available for trouble shooting the alarms