28

Calibrate Menu

This menu is used to access the various calibration procedures. It contains the following items:

ITEM

FUNCTION

Low Reference

Selecting this takes the user to the Low Reference menu

High Calibrate

Selecting this takes the user to the High Calibrate menu

Manual Entry

Shows and allows the user to set the density slope value

Clear Ref/Cal

This invokes a method that clears the density reference

Temp Ref

This invokes a method that performs the process temperature calibration

Pressure Ref

This invokes a method that performs the process pressure calibration

Auto Reference

Selecting this takes the user to the Auto Reference menu

Low Reference Menu

This menu is used to perform the low reference procedure. It contains the following items:

ITEM

FUNCTION

Reference

This invokes a method that performs the low reference procedure

Ref Density

Shows and allows the user to set the reference density value

Ref Cap

Shows the raw captured density counts

High Calibrate Menu

This menu is used to perform the high calibrate procedure. It contains the following items:

ITEM

FUNCTION

Calibrate

This invokes a method that performs the high calibrate procedure

Cal Density

Shows and allows the user to set, the calibrate density value

Cal Cap

Shows the raw captured density counts

1/uT

Shows the calculated slope value (Same value as in Calibrate Menu, Manual Entry)

Temp

Reference Menu

(This menu is ONLY used if the Hardware Menu, Proc Temp value is set for 0-10 volts or 4-20 ma.)
This menu is used to perform the process temperature reference procedure. It contains the following items:

ITEM

FUNCTION

Temp LoRef

This invokes a method that performs the low temp reference

Temp HighCal

This invokes a method that performs the high temp calibration

Temp Cal Low

Temperature value during the Temp Lo Ref procedure

Temp Cal High

Temperature value during the Temp High Cal procedure

Temp Cap Low

Shows the raw captured temperature counts during the low temp reference

Temp Cap High

Shows the raw captured temperature counts during the high temp calibration

RTD Slope

Factory value for RTD Slope

RTD Offset

Factory value for RTD Offset