Inspection, Warning, Handle padlocked off – Ronan X96S DENSITY GAGE User Manual
Page 37: Plunger padlocked off rod padlocked off, Lock tag

Density Gage Manual 012407
33
Inspection
The source holder is equipped with an ON/OFF mechanism.
During shipment and storage the
mechanism MUST BE SECURED in the OFF position
with a padlock.
Radiation Label
If the padlock is damaged, broken, or missing, contact the
RSO immediately.
Handle Padlocked OFF
General License Label
(if applicable)
Radiation Label
General License Label
(if applicable)
Plunger Padlocked OFF
Rod Padlocked OFF
Radiation Label
General License Label
(if applicable)
Lock Tag
------- WARNING -------
THIS DEVICE MAY BE MOUNTED IN PLACE
INITIALLY BY ANY PERSON PROVIDED THE
SHUTTER REMAINS LOCKED IN THE OFF
POSITION. ONLY A SPECIFICALLY LICENSED
PERSON MAY PLACE THE DEVICE IN SERVICE
BY INITIALLY OPENING THE SHUTTER AND
MAKING THE REQUIRED LEAK TEST, TESTING
FOR PROPER OPERATION OF THE ON-OFF
MECHANISM AND INDICATOR AND MAKING
THE RADIATION SURVEY.