beautypg.com

Performing a 1-point offset-only calibration, Performing a 1-point slope-only calibration, Calibrating a drop count sensor (example) – PASCO SPARKvue ver.2.0 User Manual

Page 30: Performing a 1-point offset-only calibration": 18, Performing a 1-point slope-only calibration": 18, Calibrating a drop count sensor (example)": 18

background image

Performing a 1-point offset-only calibration

In the Calibrate Sensor Enter Values screen, complete these steps:

1. Apply a known quantity to the sensor. For example, place a temperature probe in

ice water known to be 0 °C.

2. Under Calibration Point 1 Click the Standard Value box and enter the known

value.

For example, enter the known temperature of the water.

3. Under Calibration Point 1 Click Read From Sensor.

The value measured by the sensor is transferred to the Sensor Value box.

4. Click OK.

Calibration is complete.

Performing a 1-point slope-only calibration

In the Calibrate Sensor Enter Values screen,complete these steps:

1. Apply a known quantity to the sensor. For example, place a dissolved oxygen

probe in a bottle known to contain 9.1 mg/L of dissolved oxygen.

2. Under Calibration Point 2 Click the Standard Value box and enter the known

value.

For example, enter the known dissolved oxygen concentration.

3. Under Calibration Point 2 Click Read From Sensor.

The value measured by the sensor is transferred to the Sensor Value box.

4. Click OK.

Calibration is complete.

Calibrating a Drop Count sensor (example)

Tip:

The Drop Dispenser has two stopcocks. The top stop cock is used to regulate flow

rate and the bottom stopcock is used to turn the flow on and off. The bottom stopcock
should either be in the completely open or completely closed position.

The drop counter is calibrated, using the SPARKvue Experiment Tools to correlate the
number of drops counted to the volume of liquid dispensed from the Drop Dispenser.

1. Attach the Drop Dispenser (syringe, two stopcocks and drop tip) to a support rod.

2. Fill the syringe with titrant.

3. Open both stopcock valves and adjust the valve on the top stopcock so that the

drops fall at a rate of about 1 drop per second. Close the bottom valve after the rate
is achieved.

4. Connect the Drop Counter and launch SPARKvue.

Document Number: 012-11074D

18

SPARKvue

®

3   Setting up an experiment

Calibrating sensors