Clean luminex analyzer ventilation filter, Change sheath filter – Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 95

x
MAP Technology
Maintenance and Cleaning
PN 89-00002-00-150 Rev. A
7 - 7
1. Syringe Seal 2. Syringe
Figure 7-3. Front of Luminex 200 Analyzer, Access Door Opened
Clean Luminex
Analyzer Ventilation
Filter
Check the Luminex analyzer ventilation filter every 6 months.
Clean the filter only when soiled.
1. Turn off the Luminex analyzer at the power switch.
2. Unplug the power cord from the wall source.
3. On the bottom of the unit on the left side, push the filter out
toward the right side.
4. Grasp the filter and pull it out from underneath the Luminex
analyzer.
5. If the filter is dirty, clean it with a vacuum cleaner or in running,
distilled water. Stand it upright to air-dry.
6. Reinstall the filter with the arrows facing up.
Change Sheath
Filter
Change the sheath filter once a year.
1. Disconnect the sheath fluid bottle
2. Squeeze the metal quick release connections at both ends of the
filter assembly, then pull the tubing away from the filter
assembly.
1
2