Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 101

Luminex 200 Maintenance Log Month(s): ________________________________________Year:_______________
Procedure: To use this log, fill in the month(s) and year above. This form is designed to record information for four weeks. Make photocopies as necessary. Fill in the dates for which this
log is used in line one of the table below. For each item listed at the left, add your initials under each date you perform the item.
Note:
Follow your standard laboratory safety practices when cleaning or maintaining the system. Do not remove the analyzer cover under any circumstances.
Dates
Daily maintenance
Startup
Initials: (for each item listed at the left, fill in your initials under each date that you perform the item)
Laser warmup
Check sheath fluid
Check waste level
Tighten sheath cap
Alcohol Flush (70% isopropanol )
Wash twice (distilled water or
sheath fluid)
Shut down
Initials: (for each item listed at the left, fill in your initials under each date that you perform the item)
Sanitize (10%-20% bleach)
Soak (water)
Loosen sheath cap
Turn the system off (optional)
Weekly
Visual inspection
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Calibrate
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Monthly
Clean sample probe
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Wipe exterior surfaces
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Every six months
Replace air intake filter, analyzer
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Replace air intake filter, XYP
(optional)
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Replace syringe plunger seal, or
syringe
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Check analyzer ventilation filter
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Yearly
Replace sheath filter
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
As required
Replace fuses
Date/Initials:
Date/Initials:
Date/Initials:
Date/Initials:
Comments:
PN 89-00002-00-150 Rev. A