Luminex xyp specifications, Optics, Fluidics – Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 30: Electronics

Luminex 200 User Manual for LDS Version 1.7
x
MAP Technology
3 - 4
PN 89-00002-00-150 Rev. A
•
AC inlet fuse: 3 Amp, 250 V~, fast acting
Optics
•
Reporter laser: 532 nm, nominal output 10-15 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
•
Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)
•
Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm
•
Classification detector: Avalanche photo diodes with temperature
compensation
•
Doublet discrimination detector: Avalanche photo diodes with
temperature compensation
Fluidics
•
Sheath flow rate 90 µL ± 5 µL/second
•
Cuvette: 200 micron square flow channel
•
Sample injection rate: 1 µL/second ± .05 µL/second
•
Sample uptake volume: 20-200 µL
Electronics
•
Reporter channel detection: A/D resolution 14 bits
•
Communications interface: USB
Luminex XYP
Specifications
•
Ambient temperature: 15°C to 30°C (59°F to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: designed to operate at 2400 m (7874 ft.) above mean
sea level or below.
•
Physical dimensions: 44 cm (17.25 in) W x 61 cm (24 in) D x 8
cm (3 in) H
•
Weight: 15 kg (33 lbs)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 3101-1.
•
Pollution degree: UL Pollution Degree 2, as defined in section
3.7.3.2 of UL 6101A-1
•
Heater operating range: 35°C to 60°C (95°F to 140°F) with
tolerance 0°C to +2°C
•
Input voltage range: 100-240 V~ ± 10%, 0.4 Amps, 47-63 Hz
•
AC inlet fuse: 2 Amp, 250 V~, time lag
•
Communications interface: RS 232
•
Shipping and storage: allowable shipping and storage
temperature and humidity ranges, 0°C to +50°C and 20%-80%
noncondensing, respectively