Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 52

Luminex 200 User Manual for LDS Version 1.7
x
MAP Technology
5 - 6
PN 89-00002-00-150 Rev. A
7. Fill the reservoir with a solution of 70% isopropanol or 70%
ethanol. Click Retract on the toolbar to retract the plate holder.
8. Click Alcohol Flush, then click OK. The alcohol flush removes
air from the fluidics. This takes about 5 minutes.
9. Click Eject on the toolbar and replace the alcohol in the
reservoir with distilled water. Click Retract on the toolbar to
retract the plate holder.
10. Click Calibrate on the main screen. The Start Calibration dialog
box opens.
Figure 5-7. Start Calibration Dialog Box
11. Enter your name in the Operator text box.
12. Select the Product / Lot Number for the xMAP classification
calibration and reporter microspheres.
13. Click OK. The Calibration dialog box opens.
Figure 5-8. Calibration Dialog Box
14. Make sure Classification is selected, and click Start.