Materials, Preparation, Procedure – Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 128

Luminex 200 User Manual for LDS Version 1.7
x
MAP Technology
9 - 10
PN 89-00002-00-150 Rev. A
•
Micropipetters (1 µl - 1000 µl)
•
Microcentrifuge
•
Timer
•
Rotator
Materials
•
xMAP LumAvidin-modified microspheres—LIMIT
EXPOSURE TO LIGHT!
•
Biotin-conjugated molecule
•
Microcentrifuge tubes: 1.5 ml, polypropylene
•
REACTION BUFFER:
Phosphate Buffered Saline (pH 7.3 ± 0.1), Bovine Serum
Albumin (10 mg/ml)
•
BLOCKING/STORAGE BUFFER:
Phosphate Buffered Saline (pH 7.3 ± 0.1), Tween 20
(0.02% v/v), Bovine Serum Albumin (1 mg/ml), Sodium Azide
(0.05% w/v)
Preparation
1. Dilute the Biotin-conjugated molecule in REACTION BUFFER
to a concentration of 4 - 4000 nM (See Technical note 1).
Procedure
1. Centrifuge the Avidin-modified xMAP microsphere stock for 1
minute at 8,000
× g or higher.
2. Disperse the pellet with sonication, and vortex the container for
20 seconds.
3. Dispense 1.0
× 10
5
Avidin-modified xMAP microspheres into a
1.5 ml microcentrifuge tube.
4. Centrifuge the xMAP microspheres for 1 minute at 8,000
× g or
higher and remove the supernatant.
5. Add 250 µl of the REACTION BUFFER (See Technical
note 2). Vortex.
6. Add 250 µl of the Biotin-conjugated molecule solution to the
xMAP microsphere suspension. Vortex immediately.
7. Rotate the mixture for 30 minutes in the dark at room
temperature.
8. Centrifuge the microspheres for 1 minute at 8,000
× g or higher.
Aspirate the supernatant.