Replace Syringe Seal
Replace the syringe plunger seal every 6 months.
1. Turn off and unplug the Luminex analyzer.
2. Open the center door on the front of the Luminex analyzer.
3. Unscrew the knob on the syringe arm at the bottom of the
syringe, and forcefully push the syringe arm down.
4. Unscrew the syringe from the top of its housing; pull the plunger
out of the syringe.
5. Remove and replace the plunger seal.
6. Return the plunger to the syringe.
7. Screw the syringe back into the top of its housing.
8. Return the syringe arm to its original position, and hand-tighten
the screw on the syringe arm.
9. Plug in the power cord and turn the Luminex analyzer on.
10. Prime the Luminex analyzer 5 times and watch for any leaks in
the syringe area.
11. When finished, close the center door.
Warning:
The syringe arm does
not deactivate when changing
the plunger; injury could result if
the system is not unplugged.