Sample probe – Luminex 200 User Manual with LDS 1.7 Software User Manual
Page 107

x
MAP Technology
Troubleshooting
PN 89-00002-00-150 Rev. A
8 - 5
Sample Probe
Problems with the sample probe can lead to fluid leaks and
pressurization problems, as well as inhibit sample acquisition.
The sample 3-way valve is
faulty.
Contact Technical Support.
Fluid leaking from front of
analyzer
The syringe seal leaks.
Replace the syringe seal.
The syringe valve leaks.
Hand-tighten the syringe
connection (silver knob) to the
white syringe valve. Run a
Prime. If leaks continue, contact
Technical Support.
Symptom
Possible Problem
Solution
Sample probe leak
The sample probe is clogged.
Clean the sample probe. See
page 7-4.
Sample arm stuck in the up
position
The system isn’t properly
pressurized.
If a PRESSURIZATION error
message displays, see
Pressurization on page 8-3.