beautypg.com

Gf100 series – Brooks Instrument GF126 User Manual

Page 48

background image

4-4

GF100 Series

Installation and Operation Manual

X-TMF-GF100-Series-MFC-eng

Part Number: 541B137AAG

March, 2013

Section 4 Maintenance & Troubleshooting

4-5 GF100 Series Troubleshooting Guide

Symptoms & Possible Causes

Corrective Action

1. No gas flow.
Is the gas supply turned on?

Check shut-off valve and pressure readout. Open
the gas supply.

Is the regulator turned on at the

Turn off the regulator and reset it to the

correct operating pressure?

recommended pressure as described in the
Data Sheet.

Are any upstream or downstream

Verify that the valves are open and operating

shut-off valves closed, either by

properly.

the system or because of failure?
Is the MOD LED light on the GF100 Series

Observe the LED display panel on top of to

lit solid green?

verify. If the LED light is not lit, cycle power the
to reboot.

Is the commanded setpoint from

Use the tool software to verify.

tool/system at 0.00 Vdc?
Has the been commanded

Use the tool software to verify.

off by an active “valve
closed” input?
2. Flow out of range.
Is the gas inlet/outlet pressure differential

Verify that the pressure is correct for the gas

either too high or too low?

and range. If required, adjust inlet/outlet pressure
to achieve proper pressure reading.

NOTE: If the differential pressure is too high,
voltage to the will be zero, which is
abnormally low for the setpoint. If the
differential is too low, voltage to will
be at its maximum value, which is abnormally
high for the set-point.
Is the MOD LED light on the GF100 Series

Observe the LED display panel at top of .

lit solid green?

If the LED light is not lit, cycle power the to
reboot.

Is the setpoint correct for the

Use the tool software to verify.

required gas flow?
Is the calibrated for the

Check the side label. Run a

particular gas?

flow check to verify.

Is the zero correct?

Zero the according to zeroing
procedure in Section 2-14. Verify leak check rates
are OK.

3. No gas control; flow is at or
above maximum.
Is the gas pressure across the

Verify that the pressure is correct for the gas and

too high?

range. If required, adjust inlet/outlet pressure to
achieve proper pressure reading.

Are system valves open, or is the

Use tool software to verify.

purge input activated?
Is the setpoint correct for the

Use tool software to verify.

required flow?

Table 4-2 GF100 Series Troubleshooting Guide