The error/event list – B&K Precision XLN60026 - Manual User Manual
Page 65

58
characters at the end of each command when programming.
6.3
The Error/Event List
SCPI interface can offer an error/event list that contains up to 10 errors/events.
Users can read the errors/events through the “
error?” command in a first-in
first-out manner. Once an error/event is read, the read process will clear it from
the memory. To clear all errors/events from the memory, the “
*CLS” command
is used.
Error
Description
-000
No error
-001
Command error
-002
Execution error
-003
Query error
-004
Input Range error
-005
Parallel/Series function, Error mode
-006
Parallel/Series function, Multi-Master
-007
Parallel/Series function, No Slave found
-008
Parallel/Series function, Communication with Slave A
error
-009
Parallel/Series function, Communication with Slave B
error
-010
Parallel/Series function, Communication with Slave C
error
-011
Parallel/Series function, Sync. signal error when
output on
-012
Parallel/Series function, Sync. signal error when
output off