B&K Precision XLN60026 - Manual User Manual

Page 46

background image

39

After wiring is complete, configure one supply as the Master and the other
three as the Slave A, B, and C. After the power supply is configured as Master,
it will start searching for all Slaves that are connected to the Master. Therefore,
in order to setup correctly, user must set up the Slaves before
To set a XLN3640/XLN6024/XLN8018/XLN10014 to the slave mode, press

the Master is set.

,

, and

in the main screen to enter the series/parallel

setting option. Then select the SERIES mode by turning the knob (SCPI
command is “PS:MODE SERIES”) and then press

to confirm. It will

continue to the next line for the MASTER/SLAVE selection. Turn the knob to
select SLAVE A (SCPI command is “PS:TYPE SLAVEA”) for the instrument
and press

to confirm the setting. Using the same procedure to set up

SLAVE B and C for the other two supplies.

L

L

=

= L

To set the master unit, press

,

, and

in the main screen to

enter the series/parallel setting option. Then use the knob to select the series
mode (SCPI command is “PS:MODE SERIES”) and press

to confirm.

It will continue to the next line for the MASTER/SLAVE selection. Use knob
again to select MASTER (SCPI command is “PS:TYPE MASTER”) and then
press

to confirm. After confirming the supply as the Master, it will